[ECP 2026] Diagnostic Performance of the miLab™ BCM for Morphological Detection of Blasts and Immature Granulocytes in Peripheral Blood
miLab™ BCM showed close agreement with expert manual smear review and high NPV for blast and IG flags, which supports its potential as a first-line CBC and morphology screening tool.
ECP 2026 (European Congress of Pathology)
|September 12, 2026
|C.Y. Bae, et al.
Unauthorized downloading or distribution of this poster is strictly prohibited.
Supplementary Video | Reviewing AI-Flagged Cells in the miLab™ Viewer
Abstract
This study, run with Asan Medical Center, compared the miLab™ BCM with expert manual peripheral blood smear review in 361 samples. It showed close agreement across the 6-part differential and sensitive flagging of immature granulocytes (100%) and blasts (88.9%) at clinically significant thresholds.
Key Highlights
- Single-center study at Asan Medical Center (Seoul): 361 evaluable peripheral blood smear cases, compared with a certified medical technologist who has over 10 years of morphology experience.
- miLab™ BCM runs an automated CBC and AI-based WBC classification from 5 μL of whole blood in one cartridge-based run.
- Strong agreement across the 6-part differential: Passing-Bablok slopes near unity for neutrophils (1.01), lymphocytes (0.97), eosinophils (0.99) and IG (0.94).
- IG flag: 100% sensitivity, 100% NPV, ROC AUC 0.996. Blast flag: 88.9% sensitivity, 99.4% NPV, ROC AUC 0.935.
- Results reflect the AI model's standalone output before reviewer confirmation, and flagged cells can be confirmed in the on-screen gallery.
Keywords
Related Publications
[LMCE 2026] Comparative Analytical Evaluation of Automated Pre-classification by Two Digital Morphology Analyzers against Manual Microscopy for Leukocyte Differential and Abnormal Cell Detection
LMCE 2026 (Laboratory Medicine Congress & Exhibition, Seoul) | C.Y. Bae, et al. | 2026.10
miLab BCM · digital morphology · AI WBC differential · immature granulocytes
[EAPS 2026] Analytical Reliability of an AI-Powered Micro-Volume Digital Haematology Analyser across Infant, Paediatric, and Adult Specimens
EAPS 2026 (12th Congress of the European Academy of Paediatric Societies, Athens) | C.Y. Bae, et al. | 2026.09
miLab BCM · paediatric haematology · neonatal CBC · micro-volume blood test
Analytical and Morphological Performance of a Digital Image-Based Morphology Analyzer Versus a Conventional Flow-Cytometry-Based Automated Hematology Analyzer
HemaSphere | 2026;10(S1) / EHA2026 Congress | Jiyeon Kim, et al. | 2026.05
miLab BCM · digital morphology · WBC differential · blast detection